It is possible to measure oxygen concentration, carbon monoxide concentration, and combustion exhaust gas temperature. [Sato Shoji cooperates with the manufacturer (HODAKA Corporation) to support the whole country. Reliable after-sales service]
Calculations correspond to carbon dioxide concentration, combustion efficiency, exhaust gas loss, fuel-air ratio, and O2 conversion Co value.
Specifications of HODAKA combustion exhaust gas analyzer HT-1300NL
Measurement item
O 2 |
Measurement range |
0-20.9 vol.% |
accuracy |
± 0.2 vol.% |
resolution |
0.1 vol.% |
Response time |
Within 10 seconds |
CO (with H2 compensation) |
Measurement range |
0-2000ppm (maximum allowable range 4000ppm) |
accuracy |
Measured value = 0-200ppm: ± 10ppm |
Measured value = 201ppm-2000: ± 5% of measured value |
Measured value = 2001ppm ~: ± 10% of measured value |
resolution |
1ppm |
Response time |
Within 40 seconds (0-2000ppm) Within 60 seconds (2001-4000ppm) |
Exhaust gas temperature |
Measurement range |
When using a standard probe: 0 to 650 ° C |
When using a high temperature probe: 0 to 950 ° C |
accuracy |
Measured value = 0 to 100 ° C: ± 1 ° C |
Measured value = 100 ℃ ~: ± 1% of measured value |
resolution |
0.1 ℃ |
Ambient temperature |
Measurement range |
0-100 ℃ |
accuracy |
± 1 ℃ |
resolution |
0.1 ℃ |
Sensor
O 2 |
Galvanic cell |
CO (with H 2 compensation) |
Constant potential electrolysis type |
Exhaust gas temperature |
K thermocouple |
Ambient temperature |
Platinum resistance temperature detector Pt2000Ω |
Calculation item
O 2 |
0 to CO 2 Max ( calculation from O 2 ) |
O 2 conversion CO (O 2 = 0%) |
0 to calculated value ppm (O 2 is fixed at 0%) |
O 2 conversion CO (O 2 =?%) |
0 to calculated value ppm (O 2 is set%) |
Combustion air ratio |
1.0-50.0 |
Exhaust gas loss |
0-100% |
Combustion efficiency |
0-100% |
Dew point temperature |
0-100 ℃ |