Tetrachloroethylene Detection Tubes (2 to 250 ppm)Gastec Detection Tube No. 133M measures Tetrachloroethylene with a Range of 2 to 250 ppm.
- Measuring Range:
- 2 to 5 ppm
- 5 to 100 ppm
- 100 to 250 ppm
- Number of Pump Strokes: 2, 1, 1/2
- Correction Factor: 0.4, 1, 2.5
- Sampling Time: 45 seconds per pump stroke
- Detecting Limit: 0.4 ppm (2 pump strokes)
- Changes from Yellow to Reddish Purple
- Reaction Principle:
- Cl2C:CCl2 + PbO2 + H2SO4 → HCl
- HCl + Base → Chloride
- Relative Standard Deviation: 10% (for 5 to 20 ppm), 5% (for 20 to 100 ppm)
- 2 Year Shelf Life (in the refrigerator)
- Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.
- On-site measurements can be performed easily by anybody, anytime and anywhere, in a short time
- Easy-to-check direct reading possible due to calibrated scale directly printed on the tube
- An extensive measuring range can be obtained by adjusting the sampling volume
- In order to assure high precision, each production lot is tested and calibrated independently, Each detector tube has its own quality control number printed on the tube
- Long shelf-life with excellent long-term stability
TUBE NAME |
Tetrachloroethylene |
CHEMICAL FORMULA |
Cl2C:CCl2 |
MEASURING RANGE |
2-250ppm |
NO. TESTS PER BOX |
10 |
STORAGE |
Store refrigerated at 10°C (50 F) or below. |
TUBE NUMBER |
133M |