Methylcyclohexanone Detection Tubes (2 to 100 ppm)Gastec Detection Tube No.155 measures Methylcyclohexanone with a Range of 2 to 100ppm.
- Measuring Range: 2 to 50 ppm, 50 to 100 ppm
- Number of Pump Strokes: 3, 2
- Correction Factor: 1, 2
- Sampling Time: 3 minutes per pump stroke
- Detecting Limit: 0.2 ppm (3 pump strokes)
- Changes from Pale Yellow to Yellow
- Corrections for temperature & humidity: Temperature correction is necessary
- Reaction Principle: C7H12O + 2,4-Dinitrophenylhydrazine → Dinitrophenylhydrazone
- Relative Standard Deviation: 15% (for 2 to 10 ppm), 10% (for 10 to 50 ppm)
- 2 Year Shelf Life (in the refrigerator)
- Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.

- On-site measurements can be performed easily by anybody, anytime and anywhere, in a short time
- Easy-to-check direct reading possible due to calibrated scale directly printed on the tube
- An extensive measuring range can be obtained by adjusting the sampling volume
- In order to assure high precision, each production lot is tested and calibrated independently, Each detector tube has its own quality control number printed on the tube
- Long shelf-life with excellent long-term stability
CHEMICAL FORMULA |
C7H12O |
MEASURING RANGE |
2-100ppm |
NO. TESTS PER BOX |
10 |
STORAGE |
Store refrigerated at 10°C (50 F) or below. |
TUBE NAME |
Methylcyclohexanone |
TUBE NUMBER |
155 |