Isopropyl Alcohol Detection Tubes (0.02 to 5.0%)Gastec Detection Tube No. 113 measures Isopropyl Alcohol with a Range of 0.02 to 5.0 %.
-
Measuring Range:
- 0.02 to 0.04 %
- 0.04 to 2.5 %
- 2.5 to 5.0 %
-
Number of Pump Strokes: 2, 1, 1/2
-
Correction Factor: 1/2, 1, 2
-
Sampling Time: 1 minute per pump stroke
-
Detecting Limit: 0.004% (2 pump strokes)
-
Reaction Principle: CH3CH(OH)CH3 + Cr6+ + H2SO4 → Cr3+
-
Relative Standard Deviation: 15% (for 0.04 to 0.5 %), 10% (for 0.5 to 2.5 %)
-
Changes from Pale Vermilion to Pale Blue
-
Corrections for temperature & humidity: Temperature corretion is necessary
-
3 Year Shelf Life
- On-site measurements can be performed easily by anybody, anytime and anywhere, in a short time
- Easy-to-check direct reading possible due to calibrated scale directly printed on the tube
- An extensive measuring range can be obtained by adjusting the sampling volume
- In order to assure high precision, each production lot is tested and calibrated independently, Each detector tube has its own quality control number printed on the tube
- Long shelf-life with excellent long-term stability
TUBE NAME |
Isopropyl Acetate |
CHEMICAL FORMULA |
CH3CH(OH)CH3 or i-C3H7OH |
MEASURING RANGE |
0.02-5.0% |
NO. TESTS PER BOX |
10 |
STORAGE |
Store tubes in a cool dark place |
TUBE NUMBER |
113 |