Ethyl Ether Detection Tubes (10 to 1200 ppm)Gastec Detection Tube No. 161L measures Ethyl Ether with a Range of 10 to 1200 ppm.
- Measuring Range: 10 to 400 ppm, 400 to 1200 ppm
- Number of Pump Strokes: 2, 1
- Correction Factor: 1, 3
- Sampling Time: 2 minutes per pump stroke
- Detecting Limit: 2 ppm (2 pump strokes)
- Color change: Yellow → Pale blue
- Reaction Principle: (C2H5)2O + Cr6+ + H2SO4 → Cr3+
- Relative Standard Deviation: 10% (for 10 to 100 ppm), 5% (for 100 to 400 ppm)
- 2 Year Shelf Life
- Store the tubes at cool and dark place.
- On-site measurements can be performed easily by anybody, anytime and anywhere, in a short time
- Easy-to-check direct reading possible due to calibrated scale directly printed on the tube
- An extensive measuring range can be obtained by adjusting the sampling volume
- In order to assure high precision, each production lot is tested and calibrated independently, Each detector tube has its own quality control number printed on the tube
- Long shelf-life with excellent long-term stability
TUBE NAME |
Ethyl Ether |
CHEMICAL FORMULA |
(C2H5)2O |
MEASURING RANGE |
10-1200ppm |
NO. TESTS PER BOX |
10 |
STORAGE |
Store tubes in a cool dark place |
TUBE NUMBER |
161L |