Carbon Tetrachloride Detection Tubes (0.5 to 60)Nextteq Detection Tube Carbon Tetrachloride with a range of 0.5 to 60 ppm. Other substance measurable with this detector tube is Chloropicrin (See table below).
- Measuring Range: 0.5 to 2.5, 2.5 to 60
- Number of Pump Strokes: 2 to 5, 1
- Correction Factor: 1/2 to 1/5, 1
- Sampling Time: 1 minute per pump stroke
- Detecting Limit: 0.2 ppm (n=5)
- Changes from White to Yellow
- Reaction Principle:
- CCl4 + I2O5 + H2S2O7 → COCl2 COCl2 + (CH3)2NC6 H4CHO→ (CH3)2NC6 H4Cl2 (CH3)2NC6 H4Cl2 + (C6H5)2NH → Yellow product
- Coefficient of Variation: 15% (for 2.5 to 20 ppm), 10% (for 20 to 60 ppm)
- 1 Year Shelf Life
- Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.
- TLV-TWA, C (ACGIH) (ppm): .1
- Comes in a Box of 5 pre tubes and 5 detection tubes

- On-site measurements can be performed easily by anybody, anytime and anywhere, in a short time
- Easy-to-check direct reading possible due to calibrated scale directly printed on the tube
- An extensive measuring range can be obtained by adjusting the sampling volume
- In order to assure high precision, each production lot is tested and calibrated independently, Each detector tube has its own quality control number printed on the tube
- Long shelf-life with excellent long-term stability
TUBE NAME |
Carbon Tetrachloride |
CHEMICAL FORMULA |
CCl4 |
MEASURING RANGE |
0.5-60ppm |
NO. TESTS PER BOX |
5 |
STORAGE |
Store refrigerated at 10°C (50 F) or below. |
TUBE NUMBER |
134 |