Acrylonitrile Detection Tubes (0.1 to 18 ppm)This detector tube measures Acrylonitrile within 0.1 to 18.0 ppm.
- Measuring Range:
- 0.1 to 0.2 ppm - 4 pump strokes - Correction Factor: 1/2
- 0.2 to 6.0 ppm - 2 pump strokes - Correction Factor: 1
- 6.0 to 18.0 ppm - 1 pump stroke - Correction Factor: 3
- Sampling Time: 2 minutes per pump stroke
- Detecting limit: 0.05 ppm (4 pump strokes)
- Changes from Yellow to Pink
- Corrections for temperature & humidity: Unnecessary
- Relative Standard Deviation: 10 %(for 0.2 to 2.0 ppm), 5 %(for 2.0 to 6.0 ppm)
- 3 Year Shelf Life
- Reaction Principle:
-
- CH2:CHCN + Cr6+ + H2SO4 → HCN
- 2HCN + HgCl2 → 2HCl
- HCl + Base → Chloride
- Store the tubes at a cool and dark place

- On-site measurements can be performed easily by anybody, anytime and anywhere, in a short time
- Easy-to-check direct reading possible due to calibrated scale directly printed on the tube
- An extensive measuring range can be obtained by adjusting the sampling volume
- In order to assure high precision, each production lot is tested and calibrated independently, Each detector tube has its own quality control number printed on the tube
- Long shelf-life with excellent long-term stability
TUBE NAME |
Acrylonitrile |
CHEMICAL FORMULA |
CH2:CHCN |
MEASURING RANGE |
0.1-18.0ppm |
NO. TESTS PER BOX |
10 |
STORAGE |
Store tubes in a cool dark place |
TUBE NUMBER |
191L |