Acrolein Detection Tubes (3.3 to 800 ppm)Measures Acrolein with a range of 3.3 to 800 ppm.
- Measuring Range:
- 3.3 to 10 ppm - 4 pump strokes - Correction Factor 1/3
- 10 to 800 ppm - 2 pump strokes - Correction Factor 1
- Sampling Time: 2 minutes per pump stroke
- Detecting limit: 2 ppm (4 pump strokes)
- Changes from Yellow to Red
- Corrections for temperature & humidity: Temperature correction is necessary
- Relative standard deviation: 10 %(for 10 to 200 ppm), 5 %(for 200 to 800 ppm)
- 2 Year Shelf Life (in the refrigerator)
- Reaction Principle:
-
- 3CH2:CHCHO + (NH2OH)3 • H3PO4 → H3PO4
- H3PO4 + Base → Phosphate
- Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below

- On-site measurements can be performed easily by anybody, anytime and anywhere, in a short time
- Easy-to-check direct reading possible due to calibrated scale directly printed on the tube
- An extensive measuring range can be obtained by adjusting the sampling volume
- In order to assure high precision, each production lot is tested and calibrated independently, Each detector tube has its own quality control number printed on the tube
- Long shelf-life with excellent long-term stability
TUBE NAME |
Acrolein |
CHEMICAL FORMULA |
CH2:CHCHO |
MEASURING RANGE |
3.3-800ppm |
NO. TESTS PER BOX |
10 |
STORAGE |
Store refrigerated at 10°C (50 F) or below. |
TUBE NUMBER |
93 |