Anritsu AM-9000 Data collector is a high-speed multi-point data recorder containing a precision, high-speed A/D converter, a C-MOS microprocessor and large capacity memory. Multipurpose system upgrades are possible through communication with external devices through the RS-232C interface. many functions are concentrated in the compact unit and various temperature measurement regions are approached. It many collect data from as many as 12 channels!
Model |
AM-9000 |
Number of inputs |
6ch: ANP connector (Miniature Plug) |
Input connector |
ANP connector (same type thermocouple) |
Input type |
Temperature: E, K, J, T, R by selected model *2 |
Measurement range |
Temperature: TABLE Ⅰ DC Voltage: ±20.000V |
Resolution |
Temperature: 0.1℃/1℃ DC Voltage: 0.001V |
Measurement accuracy |
Temperature: TABLE Ⅱ DC Voltage : ±(0.05% of reading +0.003V) |
Memory capacity |
120,000 data |
Interface |
USB2.0 (micro-B) |
Dimensions |
92(W)×155(H)×36(D) |
Weight (Including dry batteries) |
Approx.380g |
Reference junction |
±0.3℃*3 |
Temperature coefficient |
±(0.04 × Δt ℃) *4 |
Time interval |
0.1 second to 99 minutes 59 seconds |
Signal source resistance |
500Ω or less (measurement interval 1 second or more)/100Ω or less (measurement interval less than 1 second) |
Safe excitation voltage |
Between input pins on same channels :±30VDC,Between input pins on adjacent channels : 150Vp-p |
Linearizer |
Digital Linearizer method (Compliant with JIS C1602-2015) |
Display |
Graphic liquid crystal display, 128×64 dots |
Environment operational |
Operation: 0~40℃, 15~80%RH (Non condensing) |
Battery life |
TABLE III |
Power supply |
4 AA Alkaline batteries(Supplied accessories) or AC-Adaptor(Optional accessories) |
Supplied accessories |
Manual, Test report, Softcase, Communications cable, 4 AA Alkaline batteries |
Optional accessories |
Special data analysis software, AC-Adaptor, Carrying case |
Compatible standards |
CE Marking, RoHS2 |
TABLE I: Measurement Range
Thermocouple type |
Measurement range |
E |
-200 to 800°C |
K |
-200 to 1370°C |
J |
-200 to 1100°C |
T |
-200 to 400°C |
R |
0 to 1760°C |
TABLE II: Measurement accuracy
In the environment of 25 °C ± 10 °C, when the measurement interval is 1 second or more without including the reference junction compensation accuracy.
Thermocouple type |
Accuracy compensation range |
Measurement accuracy |
E,K,J,T |
-200.0 to -100.1°C |
±1.0℃ |
-100.0 to -0.1°C |
±0.5℃ |
|
0.0 to 200.0°C |
±0.3℃ |
|
200.1 to 500.0°C |
±0.5℃ |
|
500.1℃ or more |
±0.7℃ |
|
R |
0.0℃~500.0℃ |
±1.5℃ |
500.1℃ or more |
±2.0℃ |
*If the measurement interval is less than 1 sec, add ± 0.5℃ at type E ・ K ・ J ・ T and add ± 1.0℃ at type R
TABLEⅢ 12ch model standard time for continuous use (at 25 ℃, using alkaline AA batteries)
Setting interval |
0.2sec |
1sec |
10sec |
AM-9000 Series |
approx. 50 hours |
approx. 55 hours |
approx. 140 hours |
If the number of used channels is less than 12, the usable time will be longer.