Measure oxygen from 0 to 25 vol% and notify alarm with buzzer and lamp.
Successor model of XP-3180. Equipped with Bluetooth, it can be linked with a dedicated application for smartphones and tablets. Location information and other information is automatically sent by e-mail when a gas alarm is detected. Equipped with data logging function. Dust and waterproof IP67. Intrinsically safe (Ex ia IIC T4 Ga).
Target gas: Oxygen Measurement range: 0 to 25vol%
model | XP-3380Ⅱ |
Target gas | oxygen |
Gas sampling method | automatic suction |
Detection principle | galvanic battery type |
Detection range | 0 to 25vol% |
Indication accuracy*1 | ±0.3vol% |
Display resolution | 0.1vol% |
Alarm setpoint | 18vol% |
gas alarm system | Buzzer sounds, alarm lamp (red) blinks and LCD blinks |
Error notification method | Buzzer sounds, alarm lamp (yellow) blinks and LCD blinks |
Power supply used | 4 AA alkaline batteries or 4 AA Ni-MH rechargeable batteries |
Operating temperature and humidity range | -20 to 50°C, 95% RH or less (no condensation or sudden changes in temperature and humidity) |
Working pressure range | Atmospheric pressure (800-1100hPa) |
Explosion-proof structure | Intrinsically safe explosion-proof structure (combustible gas sensor is pressure-resistant explosion-proof structure) Alkaline battery specifications: Ex ia IIC T4 Ga Ni-MH rechargeable battery specifications: Ex ia IIC T3 Ga |
Degree of protection | Equivalent to IP67 |
Continuous use time*2 | about 100 hours |
Main function | Self-diagnosis (sensor error), zero adjustment, battery level display, illumination lamp function, peak hold function, LCD backlight, buzzer stop at gas alarm, time, alarm test, buzzer volume setting, buzzer mute, data logging |
size | About W91×H164×D44mm |
mass | About 460g (including batteries) |
Standard accessories | 1m gas introduction pipe or 1m gas introduction pipe (solvent specification), shoulder belt, drain filter, filter element, alkaline dry battery (or nickel metal hydride rechargeable battery), liquid crystal protective film |
option | Leather case, alligator clip, gas inlet pipe (1, 2, 3, 5, 10m), log data collection software, bump tester, charger, attachment, LCD protective film |
*1 Based on the same measurement conditions.
*2 At 25°C, no alarm, backlight off, data logging off, Bluetooth off. It varies depending on environmental conditions, usage conditions, storage period, battery manufacturer, etc.