The TOS6200A is a tester for conducting earth continuity tests required for Class I equipment under safety standards such as IEC, EN, VDE, BS, UL, JIS, and the Electrical Appliance and Material Control Law.
Equipped with a newly developed high-efficiency power supply, it achieves a large output of 150VA while being smaller and lighter.
The constant current method eliminates the need to reset the test current even if the resistance value of the test object changes. The test time can be set from 0.3s, making it ideal for production line tests that require short takt times.
In terms of operation, we have thoroughly pursued ease of use, including a large, easy-to-read display, a memory function that stores up to 100 test conditions, and the ability to program them for automatic execution. In addition, GPIB and RS-232C interfaces are included as standard, allowing external control of test conditions such as test current, judgment resistance value, and test time, as well as readback of measured values and test results. A test lead is also included as standard, making it an attractive unit with high cost performance.
Output section |
Current setting range (*1) |
3.0~31.0A AC |
(However, if the output is below the maximum rated output and the output terminal voltage is 5.4V or below) |
||
Resolution: 0.1A |
||
Accuracy: ±(1% of setting + 0.2A) |
||
Maximum rated output |
150VA (at output terminal) |
|
strain rate |
2% or less (at 0.1Ω pure resistive load of 10A or more) |
|
frequency |
50/60Hz sine wave (selectable) |
|
Accuracy: ±200ppm |
||
Open terminal voltage |
6Vrms or less |
|
Output method |
PWM switching method |
|
output ammeter |
Measurement range |
0.0~33.0A AC |
resolution |
0.1A |
|
Accuracy |
±(1% of reading + 0.2A) |
|
response |
Average value response/effective value display (response time 200ms) |
|
Hold function |
The measured current value at the end of the test is held during the PASS and FAIL periods. |
|
output voltmeter |
Measurement range |
0.00~6.00V AC |
resolution |
0.01V |
|
Accuracy |
±(1% of reading + 0.02V) |
|
response |
Average value response/effective value display (response time 200ms) |
|
Hold function |
The measured voltage value at the end of the test is held during the PASS and FAIL periods. |
|
Resistance meter (*2) |
Measurement range |
0.001~1.200Ω |
resolution |
0.001Ω |
|
offset |
0.000~1.200Ω (with OFF function) |
|
Cancel function |
||
Accuracy |
±(2% of reading + 0.003Ω) |
|
Hold function |
The measured resistance value at the end of the test is held during the PASS period. |
|
Pass/fail judgment function |
Judgment method based on resistance value |
Window comparator method |
If a resistance value greater than the upper limit reference value is detected, it will be judged as FAIL. |
||
If a resistance value below the lower limit reference value is detected, it will be judged as FAIL. |
||
When judged as FAIL, the output is cut off and a FAIL signal is generated. |
||
If there is no abnormality after the set time has elapsed, the output will be cut off and a PASS signal will be generated. |
||
Upper standard value (UPPER) setting range |
0.001~1.200Ω |
|
Lower reference value (LOWER) setting range |
0.001~1.200Ω |
|
resolution |
0.001Ω |
|
Judgment accuracy |
±(2% of UPPER + 0.003Ω) |
|
Judgment method based on sampling voltage value |
Window comparator method |
|
If a voltage value higher than the upper reference value is detected, it will be judged as FAIL. |
||
If a voltage value below the lower limit reference value is detected, it will be determined as FAIL. |
||
When judged as FAIL, the output is cut off and a FAIL signal is generated. |
||
If there is no abnormality after the set time has elapsed, the output will be cut off and a PASS signal will be generated. |
||
Upper standard value (UPPER) setting range |
0.01~5.40V |
|
Lower reference value (LOWER) setting range |
0.01~5.40V |
|
resolution |
0.01V |
|
Judgment accuracy |
±(2% of UPPER + 0.05V) |
|
Proofreading |
Calibration with sine wave rms value using pure resistive load |
|
LED |
●PASS |
|
Lights up for about 0.2 seconds or more when judged as PASS |
||
Lights up continuously when set to PASS HOLD |
||
●UPPER FAIL |
||
Lights up when a voltage value higher than the upper limit reference value is detected and judged as FAIL. |
||
●LOWER FAIL |
||
Lights up when a voltage value below the lower limit reference value is detected and judged as FAIL. |
||
buzzer |
Buzzer ON for the set pass hold time when judged as PASS |
|
Buzzer turns on continuously in the following conditions |
||
PASS judgment when set to PASS HOLD |
||
UPPER FAIL judgment |
||
LOWER FAIL judgment |
||
The volume of the FAIL or PASS buzzer can be adjusted. However, since the settings are common, they cannot be adjusted individually. |
||
time |
test time |
Setting range: 0.3~999s |
TIMER OFF function available |
||
Accuracy: ±(100ppm of setting +20ms) |
||
environment |
Operating environment |
Indoor use, overvoltage category II |
Specification warranty range |
Temperature: 5~35℃ |
|
Humidity: 20~80%rh |
||
(However, no condensation) |
||
Operating range |
Temperature: 0~40℃ |
|
Humidity: 20~80%rh |
||
(However, no condensation) |
||
Storage range |
Temperature: -20~70℃ |
|
Humidity: 90%rh or less |
||
(However, no condensation) |
||
altitude |
Up to 2000m altitude |
|
power supply |
Input voltage range |
85~250Vac |
power consumption |
No load (READY): 60VA or less |
|
At rated load: max. 280VA |
||
Input frequency range |
47Hz~63Hz |