- Fast, new Super hybrid Sensor (SHS) gives you the ultimate in response speed - Just ONE second
- Resolution ranges: 1/60,000 to 1/600,000
- Large, bright Vacuum Fluorescent Display - see the display even in the dark
- Percentage, Counting Functions & Comparator functions
- ACAI Counting Function for highly accurate counting
- GLP / GMP / ISO Compliance

| 
Impact shock detection (ISD) provides feedback to prevent damage to the weight sensorElectronically controlled load (ECL) that enables precision management without using an external weightFlow rate display (FRD) that makes calculating and recording flow rates easier and more accurateSmart super hybrid sensor (Smart SHS) that realizes fast stabilization, long-term durability and low cost of ownershipShock logging function that records up to 50 adverse impact loads with a timestamp for later inspectionCushioning under the weighing pan to mitigate impact shocks significantlyProtection against vertical and transverse static overloadingFlow rate comparator that indicates whether the flow rate is within the designated limitsTare memory function that saves the tare value in non-volatile memoryBuilt-in fanless ionizer with Quick Ion® technology and external IR switch (GX-AE)Large and detachable, antistatic glass breeze break (0.0001 g readability models)Retractable doors that disappear into the back, requiring no extra space at the rearAutomatic self calibration in response to temperature changes or according to set schedules (GX-AE/A)Automatic minimum weight calculation with either ECL or an actual, external weightAlert function to ensure that the measured quantity is not smaller than the set minimum weightSelectable response characteristics for trading weighing speed for higher display stability or vice versaLarge, front-mounted spirit level with LED illuminationEasy-to-adjust leveling feet with thumbwheelsUser access control (UAC) that password-protects the balance from misusageGLP/GMP/GCP/ISO compliant output to meet documentation requirementsUniversal Flexi Coms (UFC) to customize printout content and layout*3Gross/Net/Tare outputData memory function to save weighing results, calibration results and unit masses to be used in counting mode in the device itselfCounting mode with the Automatic Counting Accuracy Improvement (ACAI) functionPercent mode function for target weighing or checking sample variationAnimal weighing (average & hold) function for weighing moving objectsStatistical calculation function (SCF) to facilitate the analysis of measurementsFive-level comparator function with buzzerAuto power ON and OFF functionsDensity determination function (optional density determination kits available for the 0.0001/0.001 g readability models)Underhook for weighing a magnetic material as well as measuring density | 

| 
| Models
 | GX-124AE GX-124A
 GF-124A
 | GX-224AE GX-224A
 GF-224A
 | GX-324AE GX-324A
 GF-324A
 |  
| Capacity | 122 g | 220 g | 320 g |  
| Readability | 0.0001 g |  
| Repeatability (standard deviation) | 0.0001 g | 0.0002 g (300 g) 0.0001 g (200 g)
 |  
| Minimum weight* (typical) | 120 mg |  
| Linearity | ±0.0002 g | ±0.0003 g |  
| Stabilization time (when set to FAST under a good environment) | Approx. 1.5 sec (100 g) | Approx. 2 sec (200 g) Approx. 1.5 sec (100 g)
 | Approx. 2 sec (300 g) Approx. 1.5 sec (100 g)
 |  
| Weighing pan size | Ø90 mm |  
| External dimensions | 259 (W) × 358 (D) × 332 (H) mm (including the large glass breeze break) |  
| Net weight | Approx. 7 kg |  
| Models | GF-123A | GX-203A GF-203A
 | GX-303A GF-303A
 | GX-403A GF-403A
 | GX-603A GF-603A
 | GX-1003A GF-1003A
 | GX-1603A GF-1603A
 |  
| Capacity | 122 g | 220 g | 320 g | 420 g | 620 g | 1100 g | 1620 g |  
| Readability | 0.001 g |  
| Repeatability (standard deviation) | 0.001 g | 0.002 g (1600 g) 0.001 g (1000 g)
 |  
| Minimum weight* (typical) | 1.4 g |  
| Linearity | ±0.002 g | ±0.003 g |  
| Stabilization time (when set to FAST under a good environment) | Approx. 1 sec Approx. 0.8 sec (5 g)
 | Approx. 1.5 sec Approx. 0.8 sec (5 g)
 |  
| Weighing pan size | 128 mm × 128 mm |  
| External dimensions | 212 (W) × 317 (D) × 171 (H) mm (including the small plastic breeze break) |  
| Net weight | Approx. 5 kg |  
| Models
 | GF-1202A | GX-2002A GF-2002A
 | GX-3002A GF-3002A
 | GX-4002A GF-4002A
 | GX-6002A GF-6002A
 | GX-10002A GF-10002A
 | GX-6001A GF-6001A
 | GX-10001A GF-10001A
 |  
| Capacity | 1220 g | 2200 g | 3200 g | 4200 g | 6200 g | 10200 g | 6200 g | 10200 g |  
| Readability | 0.01 g | 0.1 g |  
| Repeatability (standard deviation) | 0.01 g | 0.02 g (10000 g) 0.01 g (5000 g)
 | 0.1 g |  
| Minimum weight* (typical) | 14 g | 100 g |  
| Linearity | ±0.02 g | ±0.03 g | ±0.1 g |  
| Stabilization time (when set to FAST under a good environment) | Approx. 1 sec Approx. 0.8 sec (50 g)
 | Approx. 1.5 sec Approx. 0.8 sec (50 g)
 | Approx. 1 sec Approx. 0.8 sec (500 g)
 |  
| Weighing pan size | 165 mm × 165 mm |  
| External dimensions | 212 (W) × 317 (D) × 93 (H) mm |  
| Net weight | Approx. 5 kg |  | 