Water Vapor Detection Tubes (3 to 100 lb/MMCF)Gastec Detection Tube No. 6LP measures Water Vapor with a Range of 3 to 100 LB/MMCF.
- Measuring Range: 3 to 40 LB/MMCF, 40 to 100 LB/MMCF
- No. of Pump Strokes: 1, 1/2
- Correction Factor: 1, 2.5
- Sampling Time: 1.5 minutes per pump stroke
- Detecting Limit: 2 LB/MMCDF (1 pump stroke)
- Changes from Yellow to Greenish Purple (may become purple with high humidity)
- Corrections for temperature & humidity: Temperature correction is necessary
- Relative Standard Deviation: 10% (for 3 to 10 LB/MMCF), 5% (for 10 to 40 LB/MMCF)
- 3 Year Shelf Life
- Reaction Principle: H2O + Mg(ClO4)2 → Mg(ClO4)2・H2O
- On-site measurements can be performed easily by anybody, anytime and anywhere, in a short time
- Easy-to-check direct reading possible due to calibrated scale directly printed on the tube
- An extensive measuring range can be obtained by adjusting the sampling volume
- In order to assure high precision, each production lot is tested and calibrated independently, Each detector tube has its own quality control number printed on the tube
- Long shelf-life with excellent long-term stability
TUBE NAME |
Water Vapour |
CHEMICAL FORMULA |
H2O |
MEASURING RANGE |
3-100LB/MMCF |
NO. TESTS PER BOX |
10 |
STORAGE |
Store tubes in a cool dark place |
TUBE NUMBER |
6LP |