Vinylidene Chloride Detection Tubes (0.4 to 40.6 ppm)Gastec Detection Tube No. 130L measures Vinylidene Chloride with a Range of 0.4 to 1 ppm.
- Measuring Range:
- 0.4 to 1 ppm
- 1 to 14 ppm
- 14 to 40.6 ppm
- Number of Pump Strokes: 2, 1, 1/2
- Correction Factor: 0.4, 1, 2.9
- Sampling Time: 1.5 minutes per pump stroke
- Detecting Limit: 0.1 ppm (2 pump strokes)
- Changes from Yellow to Reddish Brown
- Corrections for temperature & humidity: Unnecessary
- 2 Year Shelf Life (in the refrigerator)
- Reaction Principle:
- CH2:CCl2 + Cr6+ + H2SO4 → HCl
- HCl + Base → Chloride
- Relative Standard Deviation: 10 % (for 1 to 4 ppm), 5 % (for 4 to 14 ppm)
- Comes in a Box of 5 pre tubes and 5 detector tubes
- Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below

- On-site measurements can be performed easily by anybody, anytime and anywhere, in a short time
- Easy-to-check direct reading possible due to calibrated scale directly printed on the tube
- An extensive measuring range can be obtained by adjusting the sampling volume
- In order to assure high precision, each production lot is tested and calibrated independently, Each detector tube has its own quality control number printed on the tube
- Long shelf-life with excellent long-term stability
TUBE NAME |
Vinylidene Chloride |
CHEMICAL FORMULA |
CH2:CCl2 |
MEASURING RANGE |
0.4-40.6ppm |
NO. TESTS PER BOX |
5 |
STORAGE |
Store refrigerated at 10°C (50 F) or below. |
TUBE NUMBER |
130L |