Ozone Detection Tubes (0.025 to 3 ppm)Gastec Detection Tube No. 18L measures Ozone with a Range of 0.025 to 6 ppm.
- Measuring Range:
- 0.025 to 0.05 ppm
- (0.05) to 0.6 ppm
- 0.6 to 3 ppm
- 3 to 6
- Number of pump strokes: 10, 5, 1, 1/2
- Correction Factor: 1/2, 1, 5, 10
- Sampling Time: 1 minute per pump stroke
- Detecting Limit: 0.01 ppm (10 pump strokes)
- Changes from Blue to White
- Corrections for temperature & humidity: Unnecessary
- Reaction Principle: 2O3 + C16H10N2O2 → 2C8H5NO2 + 2O2
- Relative Standard Deviation: 10% (for 0.05 to 0.2 ppm), 5% (for 0.2 to 0.6 ppm)
- 3 Year Shelf Life
- Store the tubes at cool and dark place.
- On-site measurements can be performed easily by anybody, anytime and anywhere, in a short time
- Easy-to-check direct reading possible due to calibrated scale directly printed on the tube
- An extensive measuring range can be obtained by adjusting the sampling volume
- In order to assure high precision, each production lot is tested and calibrated independently, Each detector tube has its own quality control number printed on the tube
- Long shelf-life with excellent long-term stability
TUBE NAME |
Ozone |
CHEMICAL FORMULA |
O3 |
MEASURING RANGE |
0.025-3ppm |
NO. TESTS PER BOX |
10 |
STORAGE |
Store tubes in a cool dark place |
TUBE NUMBER |
18L |