Aromatic Hydrocarbons Detection Tubes (120 - 0.4 to 200 ppm)Gastec Detector Tube No 120 is used for the detection and monitoring of Aromatic Hydrocarbons.
- Measuring Range:
- 0.4 to 2 ppm - 2 to 5 pump strokes
Correction Factor: 1/2 to 1/5
- 2 to 100 ppm - 1 pump stroke
Correction Factor: 1
- 100 to 200 ppm - 1/2 pump stroke
Correction Factor: 2
- Sampling Time: 1.5 minutes per pump stroke
- Detecting limit: 0.2 ppm (5 Pump Strokes)
- Changes from White to Brown
- Corrections for temperature & humidity: Unnecessary
- Relative Standard Deviation: 10% (for 2 to 20 ppm), 5% (for 20 to 100 ppm)
- 3 Year Shelf Life
- Reaction Principle: Aromatic Hydrocarbons + I2O5 + H2SO4 → I2
- On-site measurements can be performed easily by anybody, anytime and anywhere, in a short time
- Easy-to-check direct reading possible due to calibrated scale directly printed on the tube
- An extensive measuring range can be obtained by adjusting the sampling volume
- In order to assure high precision, each production lot is tested and calibrated independently, Each detector tube has its own quality control number printed on the tube
- Long shelf-life with excellent long-term stability
TUBE NAME |
Aromatic Hydrocarbons |
MEASURING RANGE |
0.4-200ppm |
NO. TESTS PER BOX |
10 |
STORAGE |
Store tubes in a cool dark place |
TUBE NUMBER |
120 |